Metrohm NIRS XDS MasterLab Analyzer User Manual

Page 162

Advertising
background image

160

▪▪▪▪▪▪▪

16

Vision reports “Auto
Linearization failed”

Auto Linearization is a program to maintain correct
wavelength position of the instrument. When this error is
reported, it may be due to one of the following reasons:

1.

In tablet transmission, the beam may be blocked.

2.

The tablet transmission reference may not be in

place.

3.

This may occur with MasterLab when connecting

from a new computer, which has no prior
knowledge of the shutdown condition of the
instrument, and has “Auto-Linearization” selected
under Configure, Options. To resolve this problem,
see Item #3 in Troubleshooting, “MasterLab
Analyzer appears to ‘lock up’ upon successful
connection.

17 Bad Voltage in Self-Test

This is related to the issue above, and to #3. The
computer has no prior knowledge of instrument settings,
and may have initialized the incorrect amplifier. Please
follow the instructions under Item #3 in Troubleshooting,
“MasterLab Analyzer appears to ‘lock up’ upon successful
connection.

If these measures do not correct the problem, please contac Metrohm service or your next local
distributor.