Metrohm NIRS XDS MasterLab Analyzer User Manual

Page 79

Advertising
background image

▪▪▪▪▪▪▪

77

When Vision first communicates with the instrument (or
after disconnection and re-connection) this screen is
shown.

Vision must establish the configuration and have it
verified by the user. This assures that test data is sent to
the correct location in the Diagnostic Database.

Click “OK to accept the instrument identification.

Vision prompts the user to insert the Tablet
Transmission Standards tray.

Click “OK” to command Vision to position the
instrument tablet transmission reference.

The user must insert the Transmission
Standards Tray into the instrument. Vision
uses these calibrated standards to perform
Reference Standardization and other key
tests.

Notice the identifying label on the right
edge of the tray – there is a mini-CD which
contains the calibrated spectra with this
same serial number.

Always use the correct file with a given
standards set. Do not substitute.

Vision scans the instrument tablet
transmission reference. Next, Vision
requests the standards file for the Tablet
Transmission Standards Set.

This file is located on the mini-CD that was
packed with the Standards Set.

Select the file and click “Open”.

This file may be loaded into the C:\Vision
directory for ease of use. The same file will
be used in other Diagnostic programs.