Metrohm NIRS XDS MasterLab Analyzer User Manual

Page 27

Advertising
background image

▪▪▪▪▪▪▪

25

13 When all the above assembly is finished, turn

on the power switch on the monochromator.
It is located on the lower surface, on the
right-hand side as shown.

The monochromator performs some
initialization tests, which take a moment.
Some noises will be heard as items find their
initial positions. This is normal.

14 Prepare to establish communication from

Vision to the XDS instrument. This is detailed
in Section 6.0, Vision Software.

This completes assembly of the MasterLab Analyzer.

Connection to the MasterLab Analyzer is covered in section 6.0.

Please be prepared to make a choice between reflectance operation or transmission operation upon
connection. Section 5.0 explains use of the instrument in various sampling modes, as a guide to this
decision.

DO NOT OPERATE OR TROUBLESHOOT WITH THE LID OPEN. SUNLIGHT MAY CONFUSE
THE OPTICAL INTERRUPTERS USED TO POSITION THE SAMPLE CARRIAGE. THIS MAY
CAUSE AN “X-Y POSITIONING ERROR” TO BE REPORTED.