Metrohm NIRS XDS MasterLab Analyzer User Manual

Page 36

Advertising
background image

34

▪▪▪▪▪▪▪

The adjustment ranges from 9.5mm to 17.25mm, in adjustment steps of 0.25mm. The effective
scanned sample areas at various spot sizes are as follows:

Spot Size

Minimum Sample Diameter.*

Scanned Sample Area

Diameter (mm)

Diameter (in.)

Square mm

Sq. Inches

9.5 mm

11.9 mm

0.47 in.

70.9 sq. mm

0.11 sq. in.

10 mm

12.5 mm

0.49 in.

78.5 sq. mm

0.13 sq. in.

11 mm

13.8 mm

0.54 in.

95.0 sq. mm

0.15 sq. in.

12 mm

15.0 mm

0.59 in.

113.1 sq. mm

0.18 sq. in.

13 mm

16.3 mm

0.64 in.

132.7 sq. mm

0.21 sq. in.

14 mm

17.5 mm

0.69 in.

153.9 sq. mm

0.25 sq. in.

15 mm

18.8 mm

0.74 in.

176.7 sq. mm

0.28 sq. in.

16 mm

20.0 mm

0.79 in.

201.1 sq. mm

0.32 sq. in.

17 mm

21.3 mm

0.84 in.

227.0 sq. mm

0.36 sq. in.

17.25 mm

21.6 mm

0.85 in.

233.7 sq. mm

0.37 sq. in.

*Based upon 125% of beam diameter


Interpolate for spot sizes in between those listed. Areas are rounded to the digits shown.