Metrohm NIRS XDS MasterLab Analyzer User Manual

Page 52

Advertising
background image

50

▪▪▪▪▪▪▪

Verify that these selections are clicked:

Instrument must stabilize before data acquisition

Use Auto-Linearization

Reference Standardization

Use Instrument Calibration

These selections are explained in more detail,
further in this manual. Because some selections do
not apply to the Tablet, they are not selected.

Reference Standardization and Instrument
Calibration must be selected for transferability
between similar instruments.

Click “OK” to exit the dialog box.

Select Configure, Tray from the menu bar.

Vision has four “default” trays built in, for ease of
use. These are common trays, and should serve
many users. A tray for Tablet Transmission
Standards is also shown. (This tray is described in
the section on Tablet Transmission.)

Check the vial diameter – or tablet placement –
that will be used, and highlight it on this list (if
shown). For example, if the tray has a 4 x 5 layout,
select the first tray on this list.

For a vial diameter not shown, click on “New”.
Vision provides a screen for creation of a new size.

The default setting of “Use Tray” is clicked. Note that the user may select “Use Iris” or Use Boat” if
another sampling method is planned. We plan to use vials, so we keep the “Use Tray” setting.

Click on “Select” to save these settings and exit this dialog box.