A new DCM screen will be shown. The “Method” name will be blank, and the “Detector” may be
marked as “Reflectance”. Proceed with these steps:

1.

Select “Transmission” for the detector. (It is highlighted in the DCM above.) This automatically

sets the “Range” to the correct setting, 800-1650nm.

2.

Enter a “Method” name. In this case we entered “MasterLab Transmission”.

Note that the “Spot Size” defaults to 9.5mm. You may hear the spot size motor running once you
accept this DCM and initiate a scan. This is normal when changing DCM types on this instrument.

Click “OK” to accept this DCM.

Now you will be asked to select the correct DCM.

The DCM created for Tablet Transmission is now
shown. Highlight it, then click on “OK”.

Vision will now connect to the instrument, in tablet
transmission mode.