To re-install the tablet transmission reference,
mount it onto the rail as shown, and screw down
the captive thumbscrew about 4-5 turns.

Do not over-tighten, or use any tools to tighten
it. The thumb screw should only be hand-tight,
and lightly so.

The thumbscrew is machined to accept a #1
Phillips screwdriver. This is for removal only, if the
thumb screw has been over-tightened.

The tablet transmission detector must be moved
to the correct position for tablet scanning. If it is
in the “stored” position, simply rotate it (gently)
90 degrees counter-clockwise, to the position
shown.

Do not force the detector beyond the intended
position. It will “click” into place.

Please verify that Vision is set for “Use Tray” in
the Select Tray Dialog box.

From the menu bar, go to Configure, Tray. In the
Select Tray Dialog box, click on “Use Tray”.

Select “Tablet Transmission Standard” as the tray
type for the tests that follow.

Always close the instrument cover when
performing tests and sample scans.