Metrohm NIRS XDS MasterLab Analyzer User Manual

Page 71

Advertising
background image

▪▪▪▪▪▪▪

69

This is a method to adjust the instrument wavelength profile to an external, traceable wavelength
standard. This feature must be used to assure method transferability.

Use Window Correction (XDS only):

This feature is used only with XDS Process instruments that operate with Transmission Pair probes.
Do not select.

Instrument matching (method transferability) requires that Reference Standardization and
Instrument Calibration be selected, and that Certified Standards are used when required to
maintain calibration.

These calibrations should be run when indicated by Photometric Test (Reference
Standardization) and Wavelength Certification (Instrument Calibration).

To initially connect, or if already connected to the
instrument in reflectance mode, use the menu as
shown. Click on Acquire, Connect.

If already connected in reflectance, Vision will
suspend that DCM connection, and will use the
new DCM to be selected.

Vision may already have one Data Collection
Method (DCM) for this instrument, as shown.

The reflectance DCM will not work in Tablet
Transmission. Instead, click on the button
marked “New”.