Metrohm NIRS XDS MasterLab Analyzer User Manual

Page 99

Advertising
background image

▪▪▪▪▪▪▪

97

Vision verifies the tray layout as
shown.

Click on “OK”.

Vision acquires a reference scan. Next, the
T02 is scanned. It is reported at right.

Vision continues with each standard in the
set.

In the upper set of spectra, the upper and
lower spectra are tolerances from the initial
IPV Setup spectrum. The IPV Setup spectrum
is the dark spectrum in the middle, displayed
in black on screen.

The lighter spectrum in the middle (red on
screen) is the current spectrum. It should be
within the upper (blue) and lower (green)
spectra as shown.

Tolerances are automatically applied, and a
“Pass” or “Fail” indication is given.