Overview of System Components - 2

Page 8

14439.C 03/2013

VerifyNow System User Manual |

USA

2.1.1

Test Device Port and Cover

The test device port holds a VerifyNow test device during the test. When seated in the port, the test

device is aligned with optical detection components to measure the agglutination of the sample using

light transmittance through wells in the test device. When closed, the test device port cover blocks

ambient light in the room from interfering with the test.

The test device port is also equipped with a spot code scanner that scans information on the test device

to determine which test is being performed and the test device lot number.

2.1.2

Display Screen

All messages and instructions for the operator display on the screen. The display screen provides

information about each step in the test, the test type, mode, as well as the current date and time. Figure

2-2 highlights the common features of the software screens.

Mode

Prompt

Test Type

Icon

26-Feb-2012

VerifyNow® P2Y12

123

14:57:00

Figure 2‑2 Display Screen

Date and Time

Feature

Description

Icons

Icons are used to navigate through the steps in a test. Icons

display along the right side of the display screen and correspond

to an adjacent icon key. Selecting an icon key displays another

screen or directs the instrument to perform an action. Refer to the

Appendix for a description of each icon.

Prompt

Most automated activities involve several steps and associated

screens. Instructions on what to do at the next step are provided

in the center of the display screen. In Figure 2-2, the image

prompts the operator to insert the sample collection tube.

Mode

The software supports two different modes, depending on

whether the instrument is performing a quality test (QC mode) or

processing a patient sample (test mode). In Figure 2-2, a patient

bed indicates a test mode.

Test Type

Displays the quality control or sample test being performed

(Aspirin, PRUTest*, or IIb/IIIa).

Date and Time

Displays the current date and time.

NOTE: The instrument goes into “stand-by” mode when not in use for more than five minutes. In stand-by mode, the screen

will be blank and the green LED indicator on the keypad will be illuminated. Pressing any key will return the instrument to

the VerifyNow instrument Main Menu. See section 8.4 to adjust this setting.