ITC VerifyNow System Platelet Reactivity Test User Manual

Page 39

Advertising
background image

Page 35

14439.C 03/2013

Patient Testing - 6

VerifyNow System User Manual |

USA

3.

From the main menu, enter

Operator ID and

Password, if required. Use the keypad to enter

the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to

enter the password number.

4.

If sufficient time has elapsed, the instrument will prompt you to perform an EQC (Figure 6-3). Refer

to Chapter 5, Quality Controls for more information.

QC

26-Feb-2012

14:57:00

NEED

EQC

Figure 6‑3 EQC Required

After a PASS has been achieved on EQC the instrument will be ready to perform a test.

5.

When the initial screen displays, press the Test

key. If the patient ID is required, the patient ID

prompt will display. If not, the instrument will prompt you to insert the test device (go to step 3).

A series of question marks will signify the required length of the patient ID established by your

institution. Use the numerical pad to input a numerical patient ID and then press the Next

key

(Figure 6-4).

Figure 6‑4 Patient ID

NOTE: A question mark is not a valid

input. All question marks must be

replaced by valid patient ID numbers in

order to proceed with performing the test.

To enter a letter for patient ID, press the Text

key.

Use the up

and down

arrows to scroll through the text option screens. At the desired

screen, press the ( ) or ( ) to move to the desired character.

Press the Text

key to select the character. It will then show up at the Text: ___ prompt.

Advance to the next character by pressing the ( ) or ( ) and repeat steps to select additional

characters.

After the Patient ID is entered, press the Next

key.

The earlier patient ID screen will display. Press the Next key again to continue. The screen will

prompt to insert the test device (Figure 6-5).

Advertising