Troubleshooting - 10

VerifyNow System User Manual |

USA

No.

Issue Description

Resolution

12

Attention - Tube or test

device removed
The sample tube or test

device was removed

during the test.

Either the sample tube/test device was removed by the operator before the test was

completed OR a non-compatible tube was used.
1) Inspect the tube, and verify that the correct Greiner tube was used. Refer to Section 6.2,

Patient Sample Collection for more information.
2) Discard the sample and the test device. Do not reuse. Re-draw a blood sample.
3) Wait for the prompt before removing the test device prior to the next test (see Section

6-14).
NOTE: Once a test has started, do not remove the test device or the sample tube

until prompted, or until after aborting the test.

13

Error - Measurement

Timeout
An error occurred during

the measurement portion

of an test or WQC.

1) Power the instrument off. Wait 10 seconds, and then power it on.
2) Perform an EQC test (see Section 5.1)
3) If the EQC is OK, run a Wet QC Level 2 (see Section 5.2.2) and verify the result falls

within expected range. If so, proceed with the test.
4) If the EQC is not OK or the WQC produces an error, call Technical Support.

14

Attention - Scheduled

EQC
An EQC test needs to be

run and current operator

does not have authority to

do so.

Typically, this event occurs when a routine user has not been given the authority to perform

a diagnostic EQC test.
1) Have an operator with QC authority return to the Main Menu screen and Login.
2) Perform an EQC test (see Section 5.1).
3) If the EQC is OK, verify the authority level of the original user (see Section 8.6), and

update, if necessary.
4) If the EQC is not OK or the problem continues, call Technical Support.

15

Attention - Failed EQC
A test was attempted

although an EQC was

required before the test

could start.

This event occurs when an test was attempted following a failed EQC, or if the date or time

was changed since the last EQC was run and an test was attempted. Typically, this event

occurs when the current user has not been given the authority to perform a diagnostic EQC

test, and the user tried to run a test when an EQC was required.
NOTE: An EQC must pass before a test can be performed.
1) Have an operator with QC authority return to the Main Menu screen and Login.
2) Perform an EQC test (see Section 5.1).
3) If the EQC is OK, verify the authority level of the original user (see Section 8.6.1), and

update, if necessary. If the EQC is not OK or the problem continues, call Technical Support.

16

Error - Mixer frequency

OR

EQC F1, F2 Parameter
The sample-mixing

rate was not within

specifications.

1) Check the alarm and error log (see Section 7.4.1) and note the frequency of Error 16 in

the log and whether it occurred during a patient test (A) or during an EQC test (E).
2) Call Technical Support with this information.
If this event occurred during an EQC test, refer to Table 10-5 for more information.

17

Error – EQC V1

parameter
This event occurs during

a Failed EQC, and is

logged as Error 17 in the

event log.

See V1 in Table 10-5.
1) Use the cleaning device (see Section 1.1) then perform an EQC (see Section 5.1).
2) Repeat if necessary. Call Technical Support if EQC does not pass after three attempts.

18

Error – EQC V2

parameter
This event occurs during

a Failed EQC, and is

logged as Error 18 in the

event log.

See V2 in Table 10-5.
1) Use the cleaning device (see Section 1.1) then perform EQC (see Section 5.1).
2) Repeat if necessary. Call Technical Support if EQC does not pass after three attempts.

19

Error – EQC P1

parameter
This event occurs during

a Failed EQC, and is

logged as Error 19 in the

event log.

See P1 in Table 10-5.
1) Use the cleaning device (see Section 1.1) then perform EQC (see Section 5.1).
2) Repeat if necessary. Call Technical Support if EQC does not pass after three attempts.