ITC VerifyNow System Platelet Reactivity Test User Manual

Page 15

Advertising
background image

Page 11

14439.C 03/2013

Overview of System Components - 2

VerifyNow System User Manual |

USA

2.3

Sample Collection Tubes

The VerifyNow System uses whole blood collected in a partial-fill vacuum tube. During the test, the tube

is inserted onto the needle in the sample well of the test device - requiring no cap removal, specimen

preparation, or pipetting step. There are several types of sample collection tubes recommended for use

with the system (Table 2-1).

Table 2‑1 Sample Collection Tubes

Collection Tube

Description

Test

Greiner Bio-One Vacuette tube
(Greiner #454322)

Partial fill, 2 mL, 3.2%
sodium citrate. Blue top

Aspirin,

PRUTest,

IIb/IIIa (only for use with ReoPro –

abciximab

)

Greiner Bio-One High Altitude
Vacuette Tube
(Greiner #454323)

Partial fill, 2 mL, 3.2%
sodium citrate. Blue top

Aspirin,

PRUTest,

IIb/IIIa (only for use with ReoPro –

abciximab

)

Greiner Bio-One Vacuette tube
(Greiner #454244)

Partial fill, 3 mL lithium
heparin. Green top

IIb/IIIa - for use with Integrilin (

eptifibatide

) or ReoPro

(

abciximab

)

Nipro
(NP-CW0185-1363080)

1.8mL citrate tube

Aspirin,

PRUTest,

NOTE: The use of sample collection tubes other than those listed may adversely affect test results.

2.4

Quality Control

Accumetrics rigorously tests each VerifyNow System to ensure it meets performance standards. In

addition, the system supports several types of quality controls to verify continued performance during

use.

2.4.1

EQC Device and Storage Slot

The Electronic Quality Control (EQC) is the primary quality control mechanism for the VerifyNow

instrument. It consists of a re-usable device that is inserted by the operator into the test port and is

used to perform a comprehensive testing routine that confirms instrument performance.

Figure 2‑6 EQC Device

When not in use, the EQC device is kept in a storage bay on the right side of the instrument.

NOTE: If the EQC device is lost or damaged, the instrument will be inoperable.