Troubleshooting - 10

Page 68

14439.C 03/2013

VerifyNow System User Manual |

USA

Electronics Diagnostics- During an EQC, the system

monitors the electronic and mechanical components, and
reports a diagnostic failure if a value is out-of-range.

The failed parameter(s) will be
indicated by an arrow (

) to the

right of the measured value(s).
Refer to Table 10-5 for solutions
organized by parameter.

CAUTION:

Seek technical assistance if you are uncertain how to

respond to a malfunction or if you are unable to correct the malfunction.

If possible, please have the numerical code that was displayed on the

screen available when contacting Technical Support.
WARNING:

The VerifyNow Instrument is not intended to be serviced by

the user. The VerifyNow instrument is to be returned to Accumetrics for

disposal or repair in the original shipping container. If there are problems

related to the VerifyNow System, contact Accumetrics Technical Support.

10.1 Messages

Table 10-2 lists alert messages and provides a brief description of the issue and instructions for

resolving it.

Table 10‑2 Troubleshooting Messages

Issue Description

Resolution

Message – Operator ID

error
The Operator ID number

entered does not match those

of registered users.

This typically occurs when the wrong operator number has been entered.
Press the Next key to return to the previous screen and re-enter your

operator identification number.

Message – Password error
The password number

entered does not match the

Operator ID.

This typically occurs when the wrong password has been entered.
Press the Next key to return to the previous screen and re-enter the

password associated with your operator identification number.

Message – Enter Patient

identification

An incorrect number of

characters for Patient ID have

been entered during a test.

This typically occurs when the patient ID number entered does not match

the specified number of characters for the patient ID.
1) Check the Patient ID displayed on the screen to see if it is the correct

length.
2) Re-enter the correct Patient ID using the correct Patient ID length. To

verify correct Patient ID length see section 8.5

Message – Retrieve patient

identification

A Patient ID has been

entered where there is no

matching record.

This typically occurs when retrieving patient test results and an incorrect

patient ID number is entered. It does not match the patient ID of the

patient tests stored in memory.
1) Check the Patient ID displayed on the screen to see if it is the correct

length.
2) Re-enter the correct Patient ID using the correct Patient ID length. To

verify correct Patient ID length, see section 8.5

Message – Remove test

device
A used test device

inadvertently remains in the

test port.

Remove the device and tube together before proceeding. The instrument

will produce two audible beeps.

Message – Perform EQC
The time between EQC

checks exceeds the set

interval specified in the

instrument settings.

The instrument prompts an operator to perform a simple electronic

quality control (EQC) check at periodic intervals.
1) Perform an EQC test (see Section 5.1).
2) If the EQC is OK, proceed with the patient test.