2 attention, error and alarm messages – ITC VerifyNow System Platelet Reactivity Test User Manual

Page 73

Advertising
background image

Page 69

14439.C 03/2013

Troubleshooting - 10

VerifyNow System User Manual |

USA

Issue Description

Resolution

Message – Non-matching

baseline values (IIb/IIIa Test

only)
The two baseline entries do

not match.

A message displays while entering the baseline PAU result if the two

entries do not match.
1) Use the Tab key to move to the field that was entered incorrectly, and

re-enter the correct baseline.
2) Press the Next key to calculate percent inhibition.

Message – Low baseline

value (IIb/IIIa Test only)
The baseline result entered

is a lower PAU value than the

post-drug result.

A flashing

and “</=” symbol will display when a suspicious baseline

has been entered.
1) Verify that the baseline result entered was correct.
2) If correct, press the Next key to calculate % inhibition as above. If

incorrect, re-enter the correct baseline as above and press the Next key

to calculate % inhibition.

Message – Reference range

baseline values (IIb/IIIa Test

only)
A baseline result is entered

that is not within the expected

range.

A reference range has been established for baseline patient samples

(before GP IIb/IIIa inhibitor administration) using the VerifyNow System.

Refer to VerifyNow IIb/IIIa package insert. This message appears if the

entered baseline is outside the reference range.
1) Verify that the baseline result entered was correct.
2) If correct, press the Next key to calculate % inhibition as above. If

incorrect, re-enter the correct baseline as above and press the Next key

to calculate % inhibition.

10.2 Attention, Error and Alarm Messages

Table 10-3 lists alerts (attention) and errors messages and provides a brief description of the issue and

instructions for resolving it.

Table 10‑3 Troubleshooting Attention and Error Messages

No.

Issue Description

Resolution

1

Attention - Bubbles in

sample
One or more of the test

detectors received too

high a light signal.

This event can be caused by an air bubble in the sample, a very low hematocrit sample, or

incomplete reconstitution of the reagent.
1) Keep the test device for troubleshooting, and verify there were no bubbles in the

detection wells. Also verify that the correct sample tube has been used. If there are

bubbles present, use a cleaning device (See Section 1.1). If there are no bubbles present,

the blood sample may have a low hematocrit. See test package insert for details.
2) Perform an EQC test (see Section 5.1). If the EQC is OK, run WQC Level 1 (see Section

5.2.1).
3) If the EQC is not OK or the problem persists, call Technical Support and have the test

device handy.

NOTE: Make sure to gently invert the sample as specified in Sections 5 and 6 to avoid

bubbles.

3

Attention - Test device

lot expired
An expired test device

has been inserted into the

instrument.

1) Check the expiration date on the packaging of the test device as well as the current date

and time on the instrument.
2) If the test device has expired, remove it and discard.

3) If the instrument’s date and time is incorrect, change the setting (see

Section 8.3).

Advertising