Procedure

1.

Locate one test device and one WQC diluent tube.

2.

Power on the instrument using the power switch on the back panel (“I” designates on). The

instrument will power on and perform a self-testing routine lasting approximately 30 seconds. After

the self-testing is complete, the Start screen will display. Press the Next key to advance to the main

menu (Figure 5-7). The instrument should be allowed to warm-up for at least 15 minutes prior to use.

26-Feb-2012

14:57:00 104A

444

QC

VerifyNow

System

®

The System is

warming up.

Figure 5‑7 Main Menu

A flashing message will display while the instrument is warming up. During this time the icon keys

are inactive and you will be unable to start a test. Go to the next step when the warm-up message

no longer displays.

3.

From the main menu, enter

Operator ID and

Password, if required. Use the keypad to enter

the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to

enter the password number.

4.

Press the QC

key. Wait for the image of the device to display.

5.

Open the foil pouch and remove the test device just before use. Hold it by the finger grip.

NOTE: Each test device has a finger grip. Avoid handling other surfaces of the test device.

6.

Remove the needle’s protective sheath by pulling directly up on the sheath (Figure 5-8). Do not twist

the sheath, as this may remove the needle.

NOTE: Do not try to re-attach any test device needles that may inadvertently removed. If this occurs, discard the device

and use a new one

Figure 5‑8 Prepare Needle