ITC VerifyNow System Platelet Reactivity Test User Manual

Page 50

Advertising
background image

Retrieving Test Results - 7

Page 46

14439.C 03/2013

VerifyNow System User Manual |

USA

In addition to viewing earlier results, icons keys enable results to be printed (optional) or another test

started.

Icon

Description

New Test

Select this icon to perform another test on the same patient

(when Patient ID is enabled). The instrument will display the

Patient ID screen with the Patient ID complete

d.

Print Result

Scroll to the selected record, and select this icon to print the

selected test result

s.

Print All

Results

Scroll backwards through results using the ( ) key, and

select this icon to print all test results in the log that are

positioned forward of the cursor, or all of the test results for a

particular patient (when Patient ID function is activat

ed).

Back

Select this icon to return to the Main Menu

.

7.2

Retrieving Quality Control

The instrument stores the last 100 EQC and WQC results in its memory. You can retrieve and print

information about the earlier EQC parameters, WQC results, test type, date and time, and operator

identification (when configured). Use the following procedure to view the quality control results.

1.

From the main menu, enter

Operator ID and

Password, if required. Use the keypad to enter

the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to

enter the password number.

2.

Press the Maintenance

key.

3.

Press the Next

key three times. Refer to the Appendix for information on navigating within the

Maintenance section.

4.

Press the EQC

key or the WQC

key to retrieve the appropriate results.

See Section 7.2.1 for a description of the EQC results log or Section 7.2.2 for a description of the WQC

results log.

7.2.1

Electronic Quality Control Results Log

The cursor is positioned on the most recent test result (Figure 7-4). Navigate within an individual EQC

test result using the ( , ) keys.

26-Feb-2012 13:51:12

EQC Results

P/F 01 02

Op V1 V2

MMDD P1 F1

HHMM F2 Em

P

0.30 1.20

00005 5.45 0.25

0429 14 14

1327 0000

F

0.30 1.20

00005 3.45 0.23

0429 0.35 14

1241 14 F080<-

S/N 1095

ALL

Figure 7‑4 EQC Results Log

EQC Result

Operator ID

Date and Time

Parameters

Flagged Parameter

Print Result

Back

Print All Results

Advertising