ITC VerifyNow System Platelet Reactivity Test User Manual

Page 79

Advertising
background image

Page 75

14439.C 03/2013

Troubleshooting - 10

VerifyNow System User Manual |

USA

Parameter

Issue Description

Resolution

V2

Maximum leak rate

This event is logged as Error 18 in the event log.
1) Use the cleaning device (see Section 1.1) then perform an

EQC (see Section 5.1).
2) Repeat if necessary.
3) Call Technical Support if EQC does not pass after three

attempts.

P1

Maximum pressure level

This event is logged as Error 19 in the event log.
1) Use the cleaning device (see Section 1.1) then perform an

EQC (see Section 5.1).
2) Repeat if necessary.
3) Call Technical Support if EQC does not pass after three

attempts.

F1

Mixer frequency flag

count

1) Check the alarm and error log (see Section 7.4.1) and note

the frequency of Error 16 in the log.
2) Call Technical Support with this information.

F2

Mixer frequency

transitions high to low

1) Visually inspect the EQC device to verify it has not been

damaged. If you have another EQC device, repeat EQC using

the second device.
2) Check the alarm and error log (see Section 7.4.1) and note

the frequency of Error 16 in the log.
3) Call Technical Support with this information.

MASK

Failing optical element

This event is logged as an error in the event log (e.g. #4, #29). It

commonly occurs when the EQC device has not been removed

completely from the test device port when prompted or when

the instrument is located in adverse lighting conditions (refer to

installation guidelines).
Call Technical Support if the EQC device had been removed

completely at the end of EQC and it fails.