Quality Controls - 5

VerifyNow System User Manual |

USA

control material does not produce a result within the expected range, perform an EQC test to ensure

that the instrument is working properly. If the EQC is OK, prepare a new WQC Level 1 sample and

repeat the WQC procedure with a new test device. If the EQC is not OK, follow the instructions in

Chapter 10, Troubleshooting. If the WQC result fails on the second attempt, contact Accumetrics

Technical Support.

17.

Press the Next

key to return to the main menu.

5.2.2

Level 2 WQC

WQC Level 2 is formulated at a clinically relevant level, and is representative of a sample with minimal

platelet inhibition. During Level 2 WQC, a pink pellet is added to the diluent to verify continued

performance, by checking that actual results are within a standard range. It takes approximately 3 to 5

minutes to complete using the following procedure:

Materials

You will need

1 test device, 1 WQC diluent tube and 1 WQC pellet

NOTE: Do not use test devices or WQC materials beyond the expiration date.

Procedure

1.

Locate one test device, one WQC pellet vial and one WQC diluent tube.

2.

Power on the instrument using the power switch on the back panel (“I” designates on). The instrument

will power on and perform a self-testing routine lasting approximately 30 seconds. After the self-testing

is complete, the Start screen will display. Press the Next key to advance to the main menu (Figure

5-18). The instrument should be allowed to warm-up for at least 15 minutes prior to use.

26-Feb-2012

14:57:00 104A

444

QC

VerifyNow

System

®

The System is

warming up.

Figure 5‑18 Main Menu

A flashing message will display while the instrument is warming up. During this time the icon keys

are inactive and you will be unable to start a test. Go to the next step after the warm-up message no

longer displays.

3.

From the main menu, enter

Operator ID and

Password, if required. Use the keypad to enter

the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to

enter the password number.

4.

Press the QC

key. Wait for the image of the device to display (Figure 5-19).