Troubleshooting - 10

Page 74

14439.C 03/2013

VerifyNow System User Manual |

USA

No.

Issue Description

Resolution

104

Alarm 104
Heater Timeout During

Heating Cycle

Power instrument off and back on. Verify the fan has at least 3 inches

of space to ventilate properly, and that it is not being affected by

airflow from a vent or another instrument.
Check if fan is functioning. If no airflow can be detected, call Technical

Support.

118

Alarm 118
Emitter problem

associated with an EQC

filter

Power instrument off and back on. If alarm occurred during EQC, use

alternate EQC device.
If no alternate EQC device is available, or alarm reoccurs, call

Technical Support.

10.4 EQC Diagnostic Messages

EQC diagnostic messages will display at the conclusion of the EQC test. Figure 10-2 provides an

example of a diagnostic message that displays when the EQC test does not produce results within the

expected range for a monitored parameter.

Figure 10‑2 Example EQC Diagnostic Messages

Look for the flagged
parameter(s) indicated
by an arrow

Identify the

parameter(s) name

1

2

To troubleshoot EQC diagnostic messages, first identify the parameter(s) flagged with an arrow, then

look up the parameter(s) in the following table. Table 10-5 lists diagnostic messages and provides a

brief description of the issue and instructions for resolution.

Table 10‑5 Troubleshooting EQC Diagnostic Messages

Parameter

Issue Description

Resolution

O1

High optical signal level

1) Visually inspect the EQC device to verify it has not been

damaged.
2) If you have another EQC device, repeat EQC using the

second device; otherwise, Call Technical Support.

O2

Low optical signal level

1) Visually inspect the EQC device to verify it has not been

damaged.
2) If you have another EQC device, repeat an EQC using the

second device; otherwise Call Technical Support.

V1

Maximum vacuum level

This event is logged as Error 17 in the event log.
1) Use the cleaning device (see Section 1.1) then perform an

EQC (see Section 5.1).
2) Repeat if necessary.
3) Call Technical Support if EQC does not pass after three

attempts.