Page 55

14439.C 03/2013

Instrument Settings - 8

VerifyNow System User Manual |

USA

8.3

Set the Date and Time

The date and time is initially configured during installation. However, this setting will need to be adjusted

during a leap year and twice annually for Daylight Savings Time. Use the following procedure to adjust

the setting:

1.

From the main menu, enter

Operator ID and

Password, if required. Use the keypad to enter

the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to

enter the password number.

2.

Press the Maintenance

key.

3.

Press the Next

key two times. Refer to the Appendix for information on navigating within the

Maintenance section.

4.

Press the Date/Time

key to display the setting (Figure 8-3).

26-Feb-2012

14:57:00

Time: 24:05
Date: 12-Jan-2001

Hour’s value must

be between 0 and

23

Figure 8‑3 Date and Time Setting

Save Setting

Current Setting

Time

Date

Back

5.

Record the new time stamp according to a 24-hour clock format. Press the Tab key to move between

the settings. For example, 1:30 pm is equivalent to 13:30 in the instrument time setting. The hour

value should be between 0 and 23. The minute value should be between 0 and 59.

6.

Record the new date stamp. Press the Tab key to move between the settings. The year value

consists of four characters and must be 1999 or later. Use either arrow key ( , ) to scroll through

the months when setting the date. The day value should be greater than 0, but not greater than the

number of days in the given month.

7.

Press the Save

key to keep the new setting and return to the Maintenance menu. Press the

Back

key three times to return to the Main Menu without saving the changes.

NOTE: In the event a setting value is not within an expected range, an error
message will display after the Save

key is pressed. Return to the setting and

correct the mistake, then press the Save

key again.

8.

Whenever an adjustment is made to the date or time setting, the instrument will require an EQC be

performed.

8.4

Set Time Out

The VerifyNow Instrument has been designed with features that restrict instrument access after a

specified time period. If the instrument is idle for a longer period, the display screen will go blank.

When an operator presses a key, the instrument automatically returns to the Main Menu and requires