Quality Controls - 5

Page 18

14439.C 03/2013

VerifyNow System User Manual |

USA

Use the following procedure to perform an EQC:

NOTE: Pressing the Back

key will stop the EQC test at any point and return to the main menu.

1.

Power on the instrument using the power switch on the back panel (“I” designates on). The

instrument will power on and perform a self-testing routine lasting approximately 30 seconds. After

the self-testing is complete, the start screen will display. Press the Next

key to advance to the

main menu (Figure 5-1). The instrument should be allowed to warm-up for at least 15 minutes prior

to use.

26-Feb-2012

14:57:00 104A

444

QC

VerifyNow

System

®

The System is

warming up.

Figure 5‑1 Main Menu

A flashing message will display while the instrument is warming up. During this time the icon keys

are inactive and you will be unable to start a test. Go to the next step when the warm-up message

no longer displays.

2.

From the main menu, enter

Operator ID and

Password, if required. Use the keypad to enter

the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to

enter the password number.

3.

Press the QC

key. Wait for the image of the device to display (Figure 5-2).

QC

26-Feb-2012

14:57:00 104A

VerifyNow

System

®

26-Feb-2012

14:57:00

Figure 5‑2 Start EQC