ITC VerifyNow System Platelet Reactivity Test User Manual

Page 96

Advertising
background image

Appendix - 11

Page 92

14439.C 03/2013

VerifyNow System User Manual |

USA

procedures or changing instrument parameters).

As an additional security feature, various instrument display screens may be set to time out after

a default time period set by Accumetrics (2 to 5 minutes), or a time period set by the user (from 0

to 255 minutes). If the instrument is idle for a longer period, it will automatically return to the Main

Menu and require re-entry of User ID if applicable.

The laboratory may decide to assign a Patient ID to each patient sample, to allow traceability of test

results. If this option is selected, the laboratory defines the format (number of characters) for the

Patient ID number and the VerifyNow Instrument will not accept a patient sample without entry of a

Patient ID in the designated format.

Electronic Quality Control: The Electronic Quality Control (EQC) is the primary quality control

mechanism for the VerifyNow Instrument. It is a re-usable device that is inserted by the operator

into the test port and is used to perform a comprehensive testing routine that confirms performance

of the following key VerifyNow Instrument subsystems.

Verifies proper performance of instrument optics.

Confirms proper functioning of the pneumatics system that draws the sample into the test device

and moves it into the test device for reaction and measurement.

Monitors reagent mixing parameters and sample data acquisition.

Confirms correct calibration parameters.

Simulates testing at two levels of results, to check correct data acquisition and calculations.

Specifically, the EQC measures two levels of turbidimetric signal that verify the dynamic range of

the instrument.

One of these signals is at the level that would be observed in a patient with

a minimal amount of platelet aggregation (Negative or Low Level Control), and the other

represents a patient who demonstrates a significant amount of aggregation (Positive or High

Level Control).

The required EQC testing frequency is input into the VerifyNow Instrument by the user. Accumetrics

recommends that the EQC be run at least on a daily basis, although the institution may select their

preferred EQC testing frequency. When the established interval has elapsed, the user is locked out

from running a patient test until the EQC test has been successfully completed.

When run according to Accumetrics’ recommendations, the EQC meets CLIA requirements

for daily QC of the instrument system.

System Controls for Each Sample Tested: Each time a test device is run on the VerifyNow System,

the instrument verifies the device expiration date, sample filling, optics performance, correct fluid

transfer, and proper mixing. The system controls prevent the operator from running an expired test

device. The system also detects certain other operator errors, such as placing the test device or the

sample in the instrument at the wrong time, or removing the test device before the test is complete.

These controls prevent reporting of an inaccurate test result. These controls exceed the CLIA

requirement for daily QC because they are performed for each test device.

11.9.1.2

VerifyNow Test Device

Test Device Internal Controls: Each test device incorporates two levels of quality control to identify

invalid test runs caused by random errors, reagent degradation or inappropriate blood samples.

Before platelet activation and fibrinogen binding begin, the Negative (Aspirin) or Low-Level (P2Y12

and IIb/IIIa) Internal Control performs a test for non-specific aggregation. During the active phase

of the test, the Positive (Aspirin) or High-Level Internal Control channel monitors the reaction and

calculates Clinical Control Units, which must fall within specified limits. A failure of either internal

control results in an error message by the VerifyNow Instrument, which prevents the reporting of an

inaccurate result.

Advertising