Cleaning and Maintenance - 9

Page 64

14439.C 03/2013

VerifyNow System User Manual |

USA

3.

Open the cover and insert the cleaning device into the test device port until it clicks (Figure 9-2). The

adhesive strip removes dust and debris on the internal pneumatic port cleaning strip.

NOTE: The instrument does not need to be in any

particular mode to use the cleaning device.

Figure 9‑2 Insert Cleaning Device

4.

Leave the cleaning device in the test port for five seconds, but never longer than ten seconds.

CAUTION:

Leaving the cleaning device in the test port for an extended

period of time can damage the instrument from adhesive adhering to the

internal pneumatic port.

5.

Remove the cleaning device completely from the test device port (Figure 9-3). Discard the device.

The cleaning device is for single use only.

Figure 9‑3 Remove Cleaning Device

6.

Repeat using another cleaning device if, upon inspection, there is visible dust and debris.

9.2

Clean Exterior Surfaces

Occasionally, a liquid may spill near the instrument or small amounts of dust may build up on the

exterior of the instrument. Periodic cleaning of the exterior surface may be performed to remove

accumulation on the instrument.

Materials

Lint-free cloth (clean room wipes)

Non-abrasive cleaner (such as Cidex

®

, pHisoHex

®

, 7X

®

, Super Edisonite

®

, 10% household bleach

solution in water, dish washing detergent, or isopropyl alcohol)