ITC VerifyNow System Platelet Reactivity Test User Manual

Page 28

Advertising
background image

Quality Controls - 5

Page 24

14439.C 03/2013

VerifyNow System User Manual |

USA

13.

When the results display, record the results or print

the results, if your instrument has an

attached printer (Figure 5-15).

26-Feb-2012

14:57:00

WQC P2Y12

Figure 5‑15 WQC Level 1 Result

14.

Open the cover. Remove the test device and tube together in one piece by grasping the test device

finger grip and pulling straight up (Figure 5-16). Never separate the tube from the test device. Close

the cover. Discard the test device and tube in a biohazard waste in accordance with institutional

guidelines.

Figure 5‑16 Remove Device

15.

At the conclusion of the test, the resulting value will be displayed on the screen. Look up the

expected values printed on the test device pouch for Control 1 (Figure 5-17).

Figure 5‑17 Interpreting WQC Level 1 Results

Control Level 1 Range

26-Feb-2012

14:57:00

WQC P2Y12

PRU:

20

16.

Compare the WQC result to the Control Level 1 range, and verify the result falls within the expected

range. If the result falls within the expected range, the instrument has passed WQC Level 1. If the