Cleaning and Maintenance - 9

VerifyNow System User Manual |

USA

Procedure

Use a recommended cleaning solution on the exterior surfaces according to the following procedure:

1.

Solutions may be applied by moistening a soft, lint-free cloth, and/or a cotton swab. Do not to allow

liquids to flow freely or be sprayed on the instrument.

2.

After cleaning with one of the solutions listed above, a cloth moistened with fresh water should be

used to dilute and remove all of the residual cleaning solution from the instrument’s surfaces.

3.

Wipe the EQC device with a damp cloth moistened with isopropyl alcohol.

CAUTION:

Do not spray or pour cleaning fluids near the test device port

or other openings. Introduction of fluids into the interior can damage the

VerifyNow instrument.
CAUTION:

Do not steam-sterilize or autoclave the instrument. Do not

immerse the instrument in any solution. Do not clean the instrument with

acetone or any other plastic solvent or abrasive cleaner.
CAUTION:

Prolonged exposure to alcohol or other strong cleansers

can result in damage or discoloration to the instrument’s case materials.

Always use a soft cloth when cleaning the display screen and bar code

window. Abrasive cleaners, pads, or cloths will permanently damage the

surface of the plastic windows.

CAUTION:

Be careful not to spill any liquids on the instrument, printer

(accessory), or power supply while cleaning the exterior surfaces.

9.3

Replace the Fan Filter

The instrument is equipped with an exterior fan vent to cool the internal electronic components.

Depending on the room conditions, small amounts of dust and debris may accumulate over time

within the fan filter and obstruct airflow. Periodic cleaning of the fan filter may be performed to avoid

overheating of the instrument.

Fan

Fan Filter

Fan Retainer

Materials

Replacement fan filter (Accumetrics Catalog No. 37056)

Replacement fan filter retainer (Accumetrics Catalog No. 37060)

Procedure

Use the following procedure to remove and clean the fan filter:

1.

Power off the instrument (“O” designates off).

2.

The fan filter is held in place by a press fit retainer. Using your fingernail or a small tool (e.g.

paperclip), gently pull from the center of the filter retainer (Figure 9-4).