Cleaning and Maintenance

The VerifyNow Instrument does not require frequent maintenance by the user. However, it is important

to perform light cleaning and simple maintenance on a routine basis in order to maintain the reliability of

the VerifyNow instrument. Table 9-1 lists a recommended cleaning schedule.

Table 9‑1 Recommended Instrument Cleaning Schedule

Inspection / Maintenance

Bi-weekly

(every other week)

Monthly

Annually

Use the cleaning device (see Section 9.1)

X

Clean exterior surfaces (see Section 9.2)

X

Replace the fan filter (see Section 9.3)

X

The recommended cleaning schedule should be used as a general guideline. Consider adjusting the

schedule based on the environmental conditions in the room.

Refer to the Preventive Maintenance

Kit Instructions for Use for specific details regarding use of the cleaning device and fan filter

replacement.

9.1

Use the Cleaning Device

Depending on the room conditions, small amounts of dust and debris may build up on the pneumatic

port connection of the instrument. A single use, disposable cleaning device is provided to remove

this debris. It consists of a clear plastic component with an adhesive strip. Use the cleaning device to

remove debris from a pneumatic port connection inside the test device port according to the following

procedure:

CAUTION:

Excessive use of cleaning devices can damage the

instrument. Cleaning more often than once a week is not recommended,

unless prompted at the display screen.

Materials

Preventive Maintenance Kit (Accumetrics Catalog No. 85062, 13/pack)

Procedure

1.

Locate the cleaning device pack, and remove cleaning device from pouch (Figure 9-1).

Figure 9‑1 Cleaning Device

NOTE: Accumetrics recommends you have at least

3 cleaning devices on hand for routine use. Refer to

the Appendix for ordering information.

2.

Grasp the cleaning device using the finger grip and

remove the opaque tape strip. Discard the

tape.