ITC VerifyNow System Platelet Reactivity Test User Manual

Page 63

Advertising
background image

Page 59

14439.C 03/2013

Instrument Settings - 8

VerifyNow System User Manual |

USA

26-Feb-2012

14:57:00

Use Password: Yes

Oper Pswd Auth

444 1323 MRQA

122 23334 RQA

Password authority

needed.

Figure 8‑6 Operator Authority Log

Save Setting

Add/Update Record

Edit Record

Enable Operator

Identification

Operator ID

Password

Authority Level

Back

Feature

Description

Enable

Operator

Identification

Displays whether operator identification is enabled. A “Yes”

indicates that operator ID and password is required to log in. A

“No” indicates this feature is not active.

Operator ID

Displays the list of numerical operator IDs currently assigned

(when configured).

Password

Displays the numerical password assigned to each user.

Authority

Level

Displays one of six authority levels assigned to each user.

Blank If no authorization level is selected, the password feature will

not activat

e

A This level allows the user to perform patient Tests and print

results

.

Q,A In addition to all features available under A, this level allows

the user to perform electronic and wet Quality control tests.

R,Q,A In addition to all features available under Q,A, this level

allows the user to Retrieve earlier patient results.

M,R,Q,A In addition to all features available under R,Q,A, this level

allows the user to change instrument settings/Maintenance.

P,M,R,Q,A

In addition to all features available under M,R,Q,A, this level

allows the user to add other users and change authority

levels and Passwords.

Icon

Description
Select this icon to save any changes made to the Operator Authority

Log and return to the prior screen.

Save

Select this icon to add an operator’s log-in, password, and authority

information to the list.

Add/Update Record

When adding or changing operator information, select this icon to add

information to the operator list.

Edit Record

Select this icon to remove the highlighted (where the cursor is

positioned) user information.

Delete Record

Select this icon to return to the Main Menu without saving the

changes.

Back

8.7

Set EQC Frequency

Electronic Quality Control (EQC) is the primary quality control mechanism for the VerifyNow Instrument.

This setting specifies the frequency at which the instrument will require an EQC test to be performed.