Reagent Storage and Handling

This section describes special considerations for receiving, storing and handling of VerifyNow reagents

and devices.

4.1

Storage and Stability

Store test devices (in their unopened pouches) and quality control materials until use at temperatures

described in Table 4-1. Do not freeze.

Table 4‑1 Recommended Storage Conditions

Test

Storage Temperature

VerifyNow Aspirin Test device

2-25ºC (36-77ºF)

VerifyNow PRUTest device

2-25ºC (36-77ºF)

VerifyNow IIb/IIIa Test device

2-8ºC – extended storage

18-25ºC (64-77ºF) - may be stored for up to 8 weeks,
but not beyond the expiration date. When removed
from refrigeration, write the discard date on the
pouches and/or kit box

VerifyNow WQC pellets and diluents

15-30ºC (59-86ºF)

Test devices and quality control materials are suitable for use until the expiration date printed on the

label. Do not use test devices or WQC materials beyond the expiration date. Refer to individual test

instructions for use, kit box and pouch for designated expiration date and recommended storage

conditions.