ITC VerifyNow System Platelet Reactivity Test User Manual

Page 61

Advertising
background image

Page 57

14439.C 03/2013

Instrument Settings - 8

VerifyNow System User Manual |

USA

8.5

Enable Patient ID

The VerifyNow Instrument supports patient result traceability. This includes a feature that assigns a

patient identification (ID) to each patient sample. The instrument’s default setting does not assign

patient ID; however, use the following procedure to adjust the setting:

1.

From the main menu, enter

Operator ID and

Password, if required. Use the keypad to enter

the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to

enter the password number.

26-Feb-2012

14:57:00

Version: S00180

Use PtId? Yes

PtId Lth:3

S/N: 895

Time out: 2 3 5

EQC Freq Daily

Use Printer? Yes

Network NONE

Figure 8‑5 Patient Identification Setting

Save Setting

Patient ID Setting

Patient ID Length

Setting

Back

2.

Press the Maintenance

key.

3.

Press the Parameter

key to display the settings (Figure 8-5). Refer to the Appendix for

information on navigating within the Maintenance section.

4.

At the Use Pt ID prompt, press either the ( ) or ( ) key to change the setting between Yes and No.

5.

If Patient ID is enabled, press the Tab key to move to “Pt ID Lth:” prompt. The patient identification

length can be any combination of letters and/or numbers up to 11 characters. Enter the number of

characters in the institution’s patient identification. For example, if a sample patient identification is

“325679”, the number “6” should be entered to designate a Patient ID length of six characters.

NOTE: After the patient identification length is defined in the system, only a Patient ID with the correct number of

characters will be accepted in order to start a test. At the start of the test, a series of “?” will represent the established

patient ID length. For example, if the ID length is set to 6 characters and the patient ID is five characters, then a leading

zero (0) must be entered. A “?” is not a valid input and all “?” must be replaced by alpha-numeric characters. If a

different number of digits is entered the

will display. The test cannot proceed until the correct number of digits is

entered.

6.

Press the Save

key to keep the new setting and return to the Maintenance menu. Press the

Back

key twice to return to the Main Menu without saving the changes.

8.6

Set Operator ID and Password

The VerifyNow Instrument has been designed with features that restrict instrument access. If this

feature is enabled, a user identification number and a matching password must be entered before

an operator can use the instrument. In addition, it is also possible to assign different authority levels

to each user, depending upon the instrument functions delegated to each user (e.g. running patient

samples, QC procedures or changing instrument settings).

The instrument’s default setting does not provide for operator identification; however, use the following

procedure to adjust the setting or add/remove users based on your institution’s policy: