VerifyNow System User Manual |

USA

9.

If the test device is the first from a new lot, the instrument will display a bar code screen as soon as

the test device is inserted into the port and the spot code is read. (Figure 6-8). Place the test device

pouch in front of the bar code scanner on the left side of the instrument so that the bar code on the

bottom edge of the pouch lines up with the scanner window. Move the pouch both towards and away

from the instrument. An audible beep will be heard when the instrument reads the information. Press

the Retry

key if you are unable to scan the bar code the first time.

VerifyNow P2Y12

U s e b a r c o d e

s c a n n e r t o r e a d

t h e b a r c o d e .

26-Feb-2012

14:57:00

Bar Code

After the bar code has been

scanned once, the instrument

will accept all remaining test

devices from that lot without

displaying the bar code screen.

Figure 6‑8 Scan Bar Code

NOTE: If you are experiencing difficulty scanning the bar code, try adjusting the angle of the scan

by raising the corner of the instrument.

10.

Gently invert the sample tube five times to mix the reagents immediately before use (Figure 6-9).

Figure 6‑9 Invert Tube

5X

11.

Wait for the Insert Tube icon to display, then insert the sample into the sample well of the test device

with the rubber stopper facing downward so that the needle pierces the stopper (Figure 6-10). The

instrument will produce two audible beeps when the tube has been fully inserted.

26-Feb-2012

VerifyNow P2Y12

14:57:00

Figure 6‑10 Insert Tube