Quality Controls - 5

VerifyNow System User Manual |

USA

Figure 5‑28 Test Progress

WQC P2Y12

26-Feb-2012

14:57:00

KEEP COVER CLOSED.

999

WQC P2Y12

26-Feb-2012

123

14:57:00

KEEP COVER CLOSED.

WQC P2Y12

26-Feb-2012

123

14:57:00

CAUTION:

The sample is pressurized during parts of the test. Never remove the

sample tube or test device during the test. To abort a test, press the Back

key.

Wait for the prompt before removing the test device and tube.

16.

When the results display, record the results or print

the results, if your instrument has an

attached printer (Figure 5-29).

NOTE: Re-draw sample if there is evidence of hemolysis or clotting in any other tube that was collected at the same time.

NOTE: Re-draw sample if there is evidence of hemolysis or clotting in any other tube that was collected at the same time.

26-Feb-2012

14:57:00

WQC P2Y12

PRU:

4

Figure 5‑29 WQC Level 2 Result

17.

Open the cover. Remove the test device and tube together in one piece by grasping the test device

finger grip and pulling straight up (Figure 5-30). Do not separate the tube from the test device. Close

the cover. Discard the test device and tube in a biohazard waste in accordance with institutional

guidelines.

Figure 5‑30 Remove Device