The test device internal controls can detect failures of the reagent system due to improper storage

or handling conditions. The internal controls will also flag an improperly collected or mishandled

blood sample, or a blood sample with certain types of interfering substances. The Test Device

Internal Controls detect errors from the reagent system, adverse environmental conditions,

and additional types of operator errors.

The Test Device Internal Controls exceed the CLIA

requirement of daily controls because two levels of control are run with each test.

Test Device Kit Temperature Indicator: Each VerifyNow Aspirin and IIb/IIIa Test kit has a

temperature indicator on the outside of the packaging. The user is instructed to inspect the indicator

upon receipt of the kit. If the indicator has changed color, the kit has been exposed to elevated

temperatures, and a Wet Quality Control (WQC) Level 2 must be run to ensure that the reagents

are performing properly. The Temperature Indicator detects errors due to adverse environmental

conditions.

11.9.1.3

VerifyNow Assay WQC

Wet Quality Controls Levels 1 and 2 are available from Accumetrics for verifying the integrity of the

VerifyNow System and for evaluating operator proficiency. WQC Level 1 and Level 2 are formulated

at clinically relevant levels. In the VerifyNow Tests, Level 1 is representative of a sample with

platelet inhibition, and Level 2 is representative of a sample with no platelet inhibition.

Laboratories may elect to run these controls on a regular basis for QC of the VerifyNow

Tests, although Accumetrics believes that the comprehensive QC components built into every

VerifyNow test make this unnecessary. Accumetrics recommends that Level 2 WQC be run

under the following conditions:

Whenever a new shipment or lot number of VerifyNow Test Devices is received.

When a temperature indicator changes color, indicating that a VerifyNow Aspirin or IIb/IIIa Test kit

has been exposed to elevated temperatures.

In the case of repeated failures of the Test Device Internal Control, to rule out degradation of test

device reagents as the cause of the failure.

VerifyNow Assay WQC kit (part number 85047) may be used as a tool for other types of activities

required by the CLIA Regulation, other than the daily quality control requirements. These activities

include the following:

Proficiency testing

Laboratory personnel competency evaluation

Establishment and verification of system performance specifications (493.1253)

Analytical system quality assessment (493.1289)

Accumetrics Technical Support can advise on the use of WQC as tools in meeting these CLIA

requirements.

NOTE: The recommendation for the VerifyNow IIb/IIIa Test only is to run Level 2

WQC with each lot or shipment, or every 30 days.

11.9.1.4

Accumetrics Quality Control Procedures

To assure that VerifyNow System components are of the highest quality, Accumetrics subjects its

products to rigorous quality control procedures before they are shipped to customers.

Each VerifyNow Instrument is tested mechanically, optically and electrically throughout the

manufacturing process. The instrument’s performance is validated by comparing it to Accumetrics’

internal reference standards. The validation process consists of the following: