Quality Controls - 5

Page 28

14439.C 03/2013

VerifyNow System User Manual |

USA

12.

Replace the cap on the diluent tube by pressing and turning simultaneously. Immediately after

replacing the cap, gently invert the tube five times to mix the reagents (Figure 5-25). Control material

should be used within 15 minutes of reconstitution.

Figure 5‑25 Invert Sample

5X

If the reconstituted control is not used immediately, gently invert the tube five times to mix the

reagents again just before use (Figure 5-26).

13.

Wait for the Insert Tube image to display, then insert the WQC sample into the sample well of the

test device with the rubber stopper facing downward so that the needle pierces the stopper (Figure

5-26). The instrument will produce two audible beeps when the tube has been fully inserted.

26-Feb-2012

WQC P2Y12

14:57:00

Figure 5‑26 Insert Tube

14.

Close the cover to the test port (Figure 5-27).

Figure 5‑27 Close Cover

15.

The instrument automatically draws the WQC sample from the vacuum collection tube into the test

device, heats the sample, and proceeds with the analysis of the sample. During this time the screen

will flash an icon of the test device indicating that the sample is processing. Do not open the cover

until the test has completed and a result displays. A calculator will display when the test is near

completion (Figure 5-28).

P2Y12