ITC VerifyNow System Platelet Reactivity Test User Manual

Page 74

Advertising
background image

Troubleshooting - 10

Page 70

14439.C 03/2013

VerifyNow System User Manual |

USA

No.

Issue Description

Resolution

4

Error – Test device spot

code
The instrument does not

recognize the spot code

on the test device.

This error may occur if the instrument is located under bright light or if the spot code on

the test device does not match the bar code scanned from the test device pouch (lot

information).
Place the instrument away from direct sunlight, bright light, or other radiant heat sources

(refer to installation guidelines).
1) If the error occurs when running an test, check to see if the spot code label is marred or

crooked. If so, use another test device.
2) If this error occurs while attempting to run an EQC, turn the instrument on and off. Next,

perform another EQC (see Section 5.1).
3) Scan the bar code from another pouch from the same lot (do not open the pouch).
4) If the problem continues, call Technical Support.

5

Error - Invalid bar code
The scanned bar code

label is not the proper

length or type

1) Press the Back

key, and scan a bar code from another pouch from the same lot (do

not open the pouch).
2) If this does not work, perform an EQC test to ensure the instrument is working properly

(see Section 5.1). If EQC is OK, try the test again.
3) If the EQC is not OK or the problem continues, call Technical Support.

6

Attention – Printer not

responding
There is a problem with

the printer that prevents

printing.

The printer is not responding. The printer may be turned off or the printer not properly

attached to the instrument.
1) Check to make sure the printer is turned on. If not, turn the printer on and press the

Repeat

key to start printing.

2) Ensure that the printer and instrument are correctly connected and that the printer cable

is seated correctly at the back of the instrument.
3) If the problem continues, call Technical Support.

7

Attention – Printer out

of paper
There is a problem

with the printer roll that

prevents printing.

The printer may be out of paper or there may be a paper jam.
1) Make sure the printer is loaded with paper. If not, insert paper into the printer and press

the Repeat

key to start printing.

2) Check that paper moves properly through the printer. Press Repeat

key to start

printing.
3) If the problem continues, call Technical Support.

8

Attention – Printer not

ready
There is a problem with

the printer that prevents

printing.

The printer is not responding. The printer may be turned off or the printer not properly

attached to the instrument.
1) Turn the printer on and off. Then press the Repeat key to start printing.
2) Ensure that the printer and instrument are correctly connected and that the printer cable

is seated correctly at the back of the instrument.
3) If the problem continues, call Technical Support.

10

Attention - Detection

wells fill
The sample did not fill

the detection wells in

time, which is required for

consistent activation of

platelets.

Typical causes of this event include a low or high fill volume in the sample tube, wrong

sample tube use, or not inserting the sample tube firmly against needle hub.
1) Keep the test device, and verify there was adequate volume of blood and the correct

sample tube was used. Refer to Section 6.2, Patient Sample Collection for more

information.
2) Use the cleaning device (see Section 9.1).
3) Perform an EQC test (see Section 5.1)
4) If the EQC is OK, run a Wet QC Level 1 (see Section 5.2.1) and verify the result falls

within expected range.
5) If the EQC is not OK, call Technical Support and have the test device handy.

11

Attention - Air leak
The sample left the

staging well prematurely

during the warming cycle.

Typical causes of this event include a dirty port connection, use of an incompatible sample

tube, and not inserting the sample tube firmly against needle hub.
1) Keep the test device for troubleshooting, and verify the detection wells filled with blood

and the correct Greiner tube was used. Refer to Section 6.2, Patient Sample Collection for

more information.
2) Use the cleaning device (see Section 1.1).
3) Perform an EQC test (see Section 5.1)
4) If the EQC is OK, run a Wet QC Level 1 (see Section 5.2.1) and verify the result falls

within expected range.
5) If the EQC is not OK, call Technical Support and have the test device handy.

Advertising