ITC VerifyNow System Platelet Reactivity Test User Manual

Page 26

Advertising
background image

Quality Controls - 5

Page 22

14439.C 03/2013

VerifyNow System User Manual |

USA

7.

Open the cover. Using the finger grip, insert the test device into the test device port until it clicks

(Figure 5-9).

Barcode

Scanner

Figure 5‑9 Insert Device

8.

If the test device is the first from a new lot, the instrument will display a bar code screen as soon as

the test device is inserted into the port and the device spot code is read. (Figure 5-10). Position the

barcode of the test device pouch in front of the bar code scanner on the left side of the instrument

so that the bar code on the bottom edge of the pouch lines up with the scanner window. Move the

pouch both towards and away from the red barcode light. An audible beep will be heard when the

instrument reads the information. Press the Retry

key if you are unable to scan the bar code the

first time.

Figure 5‑10 Scan Bar Code

Bar Code

WQC P2Y12

U s e b a r c o d e

s c a n n e r t o r e a d

t h e b a r c o d e .

26-Feb-2012

14:57:00

After the bar code has been scanned once, the instrument will accept all remaining test devices from

that lot without displaying the bar code screen.

NOTE: If you are experiencing difficulty with the bar code scanner scanning the label, try adjusting the angle of the scan by

raising the corner of the instrument.

9.

Gently invert the diluent tube five times to mix the reagents immediately before use (Figure 5-11).

Figure 5‑11 Invert Tube

5X