Rechargeable batteries may be supplied by Casella CEL in a discharged state.

In order to obtain maximum capacity from new NiCd batteries, several charge

and discharge cycles may be required.

Serial Number Label

Boot Mode
Normally
Disabled

Charger
Enable

Factory
Settings

00014

ON

1

2

Figure 11: Location of DIP switches

Warning !

DO NOT enable the charger circuit while non-rechargeable

batteries are installed.
The instrument has a built in

fast-charging circuit that uses

power applied via the 12 V DC

socket shown in Figure 8. This

allows rechargeable batteries to

be recharged in-situ. When

rechargeable batteries are to be

charged in situ, set the Charger

Enable DIP switch to ON as

shown in Figure 11.

Charging will terminate automatically once the batteries are fully charged,

which takes approxi- mately 3 hours for fully discharged batteries.

2.1.2 Sampling Probe*

The detachable probe houses the measurement chamber together with the

associated optics, infra red emitter and receiver. Connect the probe by

inserting the cable connector into the socket in the top end of the instrument

case, with the red dot facing the front of the instrument to ensure correct pin

location.
DO NOT pull on the cable to disconnect the probe.
Disconnect the probe and cable from the instrument ONLY by pulling out-

wards on the knurled sleeve, in the direction of the arrows shown in Figure 4.

The probe may be un-clipped from the control unit for hand held operation.

2.1.3 Supporting the Instrument

For general monitoring applications the instrument may be fitted to a suitable

tripod (available from Casella CEL).

Note*:

Reading errors may be observed when using the instrument in very bright

or changing ambient light conditions because the measurement method

depends on the scattering of infra-red light. Therefore, exercise care when

using in direct sunlight. Shield the probe from direct light sources and hold

it steady while taking a reading. An aspirated or PUFF adaptor may be used

to minimise light ingress.

Page 14 of 60