Command ‘Z’: - Stop real time output

E.g. ‘Z’ ; This will inhibit the real-time data output.

Command ‘I’: - Interrogate target

E.g. ‘I’
The instrument will transmit a comma-delimited string containing two ID

strings. The first is the number of records stored in the loggers and the

second is the firmware version of the instrument.
i.e.

“CasellaCEL,(01234) 844100,K, 11,80-176087A-XX,[CR][LF]”

(This information can be viewed in the status screen of the instrument.)

Command ‘T’: - Download records stored

E.g. ‘T’ ; Download all records stored in the instrument
i.e.

Log close, 28/07/03,12:13:48, 129.396, 129.445,28/07/03,12:13:42,

0 [CR][LF]

28/07/03,12:13:48, 129.361, 1 [CR][LF]

28/07/03,12:13:46, 129.387, 2 [CR][LF]

28/07/03,12:13:44, 129.439, 3 [CR][LF]

Log close, 28/07/03,12:11:30, 240.920, 240.947,28/07/03,12:11:22,

4 [CR][LF]

28/07/03,12:11:30, 240.882, 5 [CR][LF]

28/07/03,12:11:28, 240.909, 6 [CR][LF]

28/07/03,12:11:26, 240.897, 7 [CR][LF]

28/07/03,12:11:24, 240.920, 8 [CR][LF]

28/07/03,12:11:22, 240.937, 9 [CR][LF]

28/07/03,12:11:20, 240.935, 10 [CR][LF]

28/07/03,12:11:18, 240.943, 11 [CR][LF]

28/07/03,12:11:16, 240.935, 12 [CR][LF]

Command ‘R’: - Set instrument id and time date

This command will enable the user to remotely set time/date and two ID

string. All records stored in the instrument will be erased.
(This information can be view in the status screen of the instrument).
E.g. ‘R’ ; wait until ‘D’ received then transmit data.
“CasellaCEL,(01234)844100,K,28/07/03,12:30[CR][LF]”
NOTE: ID max string length 16 bytes each. Date format is dd/mm/yy. Time

format is hh:mm in 24 hour format.