Physical

Display:

128 x 64 pixel LCD Graphics panel with

backlight.

Keypad:

7 key tactile membrane

Weight:

Instrument with probe: 0.97 kg.

Complete kit in carry case: 4.5 kg

Dimensions:

Instrument H x W x D: 245 x 95 x 50 mm.

Probe: 35 mm ∅ x 290 mm total length.

Carrying case H x W x D: 135 x 490 x 370

mm.

Displayed Values

Current Concentration

Represents the rolling average of instan-

taneous dust concentration over a user

selectable time period from 1 to 60 sec.

Time weighted average

Representing the average value since

switch on or reset.

Max Value

Representing the maximum instantaneous

value since switch on or reset.

(Instantaneous values are updated every

second)

Scrolling Graph

Graph mode provides unique indication of

variations in dust concentration.

Selectable timebase periods of 100 sec,

200 sec, 15 mins, 60 mins,

Y-axis range may be autoranging or fixed.

Factory calibration

Factory calibrated using traceable

isokinetic techniques and ISO 12103-1 A2

Fine test dust (Arizona Road Dust

equivalent).

Four user defined calibration settings to

suit specific dust types or monitoring

applications.

Routine Calibration

Easy software calibration for Zero and

Span setting.

Confirmation of Factory calibration point

using optical reference filter.

Data Logger

Internal memory:

64 k EEPROM providing over 15,700 data

points.

Example capacities:

Logging every 5 seconds: 21 hours

48 mins,

Logging every minute: 10 days 21 hours

Logger Interval:

Adjustable from 2 to 600 seconds.

Page 52 of 60

Specification

MICRODUST pro Aerosol Monitoring System &

WINDUST pro Application Software - Handbook