Casella CEL Software for the Microdust dust monitor User Manual

Page 19

Advertising
background image

Fast response shows sudden changes in

dust concentration.

Average:

The analogue output is derived from the

averaging time set in the instrument and

provides a smoother instrument response.

Alarm:

The output voltage will represent a set

digital value where 2.5 V is Alarm On and

0 V is Alarm Off.

When Alarm is selected, the following additional options come into

operation.
Status:

On enables the alarm,

Off disables the alarm.

Alarm Level:

Allows an alarm trigger level be set as

xxx.xx (mg/m

!

or mg/l). This mass

concentration level is taken from the

average reading selected in this menu.

When the concentration level is equal to

or greater than the alarm level, the alarm

is triggered instantaneously.

Alarm Reset:

Auto resets the alarm when the

concentration level has fallen below the

alarm level for more the 5 seconds.

Keypad resets the alarm when any key is

pressed.

18.

When all required settings for the proposed task have been selected,

press

to accept them and return to the Configuration menu.

19.

Specify the display back light function by pressing

while high-

lighting the Back Light option.

20.

Choose from the following options:
Light ON

The display is lit as you leave this screen.

It remains lit until switched off.

Light OFF

Switches the backlight off.

Keypress ON

The display is lit each time a key is

pressed while viewing a menu.

It remains lit for the duration set.

Duration

Sets the duration timer.

The duration may be set in 1 second steps anywhere between 5 and

30 seconds.

21.

Press

to accept the setting and return to the Configuration menu.

Enter

Enter

Enter

Page 19 of 60

Operation

MICRODUST pro Aerosol Monitoring System &
WINDUST pro Application Software - Handbook

Advertising