3 fault finding tips – Casella CEL Software for the Microdust dust monitor User Manual

Page 50

Advertising
background image

If this is not done, the clean purge air emerging from the measurement

chamber will prevent entry of particulate matter.

5.2.2 Cleaning

When it is suspected that lens contamination has occurred (recognised by

the inability to zero or calibrate the instrument correctly), attach the air purge

bellows to the probe inlet and pump vigorously with the measurement

chamber open. Any dust deposited on the surface of optical components

should be blown clear.
To prolong the serviceable life of the instrument, it is recommended that this

operation be performed after each use.
If the lenses become so heavily contaminated that using the air purge pump

is ineffective, the instrument should be returned to Casella CEL for cleaning

and re-calibration. Note that the Warranty does not extend to cleaning or

general servicing of the instrument.

5.3

Fault Finding Tips

The MicroDust pro contains no user serviceable components. Repairs should

be performed only by authorised personnel.
If an electrical fault is found, or a fault persists after carrying

out the recommendations in the service hint, return the

instrument to Casella CEL Ltd for repair.

Fault Symptom

Possible Cause

Service Hint

Instrument does not

switch ON

Battery discharged, or

failing to charge. Confirm

charger indicator is lit

Recharge or replace

battery if necessary.

Use external power supply.

Probe always reads zero

Fault with Infrared

transmitter or receiver.

Check the probe

connection.

Continuous “over range”

indication

Optical misalignment due

to mechanical shock.

Excessive contamination.

Return the instrument to

the service department for

optical alignment, cleaning

and re-calibration.

Unstable or drifting zero in-

dication with probe cover

on

Sample chamber may be

contaminated by dust or

fibre.

Use the bellows pump to

purge the probe with clean

air prior to setting zero.

The probe cover should be

open slightly when

purging, to allow venting of

purge air.

Probe contamination

message is displayed

Excessive contamination of

probe or optical

misalignment due to

mechanical shock.

Purge the probe with clean

air from the bellows pump.

Page 50 of 60

Maintenance
& Servicing

MICRODUST pro Aerosol Monitoring System &

WinDust pro Application Software - Handbook