Casella CEL Software for the Microdust dust monitor User Manual

Page 23

Advertising
background image

Each probe is supplied with its own calibration filter that creates a

fixed optical scattering effect within the sampling space. This fixed

reference can be used to confirm the factory calibration point for the

instrument and probe.
Use only the optical filter supplied for use with the probe.
The reference value for the calibration filter supplied for the probe is

shown on the filter and is also displayed at the bottom of the Set

Span screen.

10.

Insert the calibration filter into the measurement chamber with the

arrow pointing towards the probe handle.

11.

Press

and wait a few seconds for the reading to stabilise.

The display should give a reading within ±1 digit of the value shown

on the calibration filter to confirm that the factory set sensitivity of

the instrument has not been changed.

12.

If the correct value is not displayed (for example because the instru-

ment was last used with a different probe and filter) display the Set

Optical Filter screen.

13.

Press

and edit the entries to match the information given on the

filter.

14.

Press

to accept the changes.

This is an absolute calibration of the instrument that remains valid

when a correction factor is applied. After calibration, if the probe

becomes heavily contaminated, a warning will be displayed and the

contaminated ranges disabled.

2.3.2 Calibrating for a Specific Dust Type

All light scattering instruments exhibit some sensitivity to a particulate’s

refractive index, size, shape and colour. When faced with varying dust types,

the sensitivity of the instrument will differ slightly due to variations in

particulate size, refractive index and colour. The MicroDust pro‘s use of a

narrow forward scattering angle minimises these effects but for optimum

calibration to a specific dust type, gravimetric calibration may be required.
The optional gravimetric and respirable adapters incorporate 25 or 37 mm

filter cassettes and provide a convenient method of gravimetrically calibrating

MicroDust pro instruments. Standard gravimetric techniques can be used

with the modules to give a cross check between the readings of the

MicroDust pro and a weighed sample. This is achieved by collecting two

Enter

Enter

Enter

Page 23 of 60

Operation

MICRODUST pro Aerosol Monitoring System &
WINDUST pro Application Software - Handbook

Advertising