2 tsp / respirable adapters – Casella CEL Software for the Microdust dust monitor User Manual

Page 44

Advertising
background image

4.2

TSP / Respirable Adapters

The basic design of the TSP and respirable adapters is the same, but with

different inlet fittings. The respirable adapter uses the Cyclone principle of

dust separation to draw only the respirable fraction of particulate matter

through the measurement chamber and on to the filter (D50 cut = 4 µm at

2.2 l/min).
The cyclone is fitted to the mounting block using an “O” ring seal. A suitable

constant flow pump set for 2.2 l/min such as the Casella Apex, Apex pro,

Vortex Standard 2, or Timer 2 should be connected to the filter cassette

assembly using silicone tubing.
The cyclone must operate in the upright position and the “coarse dust

container” (rubber grit pot) should be cleaned periodically and checked for

damage.

Pump Connection

ø37 mm Filter Holder

00026

"O" Ring

"O" Ring

Adapter

Locating Spigot

Inlet Tube Complete

"O" Ring

Thumb Screw

ø25 mm
Filter Holder

ø37 mm
to ø25 mm
Reducer

"O" Ring

Measuring Chamber

MicroDust Probe

"O" Ring

Coarse Dust
Container

Locating Spigot

Cyclone Complete

Figure 19: Arrangement of Gravimetric and Respirable adapter modules

Page 44 of 60

Sampling
Accessories

MICRODUST pro Aerosol Monitoring System &

WINDUST pro Application Software - Handbook