5.

MAINTENANCE & SERVICING

5.1

Service Department

Casella CEL’s in house service department offers a comprehensive range of

repair and calibration services, designed to effect a fast and efficient back-up

for all our products. The Service Department is operated under the scope of

our BSI registration for products manufactured by us. We will however,

undertake the repair of other manufacturers equipment. For further

information please contact our service department at our Bedford

headquarters. We will be happy to provide quotations for individual repairs or

provide annual maintenance under contract.
We recommend factory service by technicians trained and equipped to repair

your instrumentation. Should you wish factory repair assistance, send your

equipment in a package equivalent to the original packaging. Insure to full

value and ship pre-paid. Include a letter giving full details with your packing

list.

Send to:

Casella CEL Ltd

(Service Department)

Regent House

Wolseley Road

Kempston

Beds MK42 7JY

United Kingdom

For service outside the United Kingdom, please return to our appointed

distributor.

5.2

Lens Contamination

As with all optical dust measurement equipment, lenses may eventually

become contaminated with particulate that has settled on the surface. In the

long term, this can reduce the instrument’s reading for a given dust

concentration or cause difficulty in achieving a stable zero indication. By

careful design of the optics and measurement chamber, the need for regular

cleaning of the lenses has been greatly reduced.

5.2.1 Long Term Monitoring

For long term sampling or where high contamination levels are anticipated,

continuous purging by a source of dry filtered air is recommended. Use one

of the pumps from the Vortex series. The flow rate to the purge inlet should

be maintained at approximately 10% of the main sample rate.
When continuous purging is adopted, it is necessary to draw the aerosol

through the probe’s measurement chamber using an aspiration technique.

Page 49 of 60

Maintenance

& Servicing

MICRODUST pro Aerosol Monitoring System &
WinDust pro Application Software - Handbook