Casella CEL Software for the Microdust dust monitor User Manual

Page 41

Advertising
background image

The gravimetric correction function automatically compares a true gravimetric

concentration of dust with the corresponding data stored in the PC date file.
1.

Load the relevant data file, as described in Section 3.5, ensuring that

it corresponds with the gravimetric result.
The data summary will be displayed.

2.

Select

Calibration

from the

Edit

menu or click

.

A dialog will be opened.
If the file has already been corrected for gravimetric results, a warning

will be given. (To remove an existing correction refer to Step 7.)

3.

Enter the gravimetric filter result in the Gravimetric Sample field (i.e.

the gravimetric density determined as in Section 2.3.2).

4.

When required, enter information in the relevant field to identify the

calibration for the file.

5.

Select

OK

to proceed.

The software compares the gravimetric result with the measured

average from the file and calculates a gravimetric correction factor.
When a graph or table is displayed for this file, the correction factor

will be applied automatically to the data so that corrected results are

displayed.
When the corrected graph or table is closed, you will be asked to

save the changes in the file.

6.

If required, select

Yes

.

The modified file will be saved with the original file name and

extension (

.dat

).

7.

If you want to remove any corrections applied to a datafile, activate

the correction function and enter “0" in the Gravimetric Sample field.
The existing corrections will be removed.

Page 41 of 60

Application

Software

MICRODUST pro Aerosol Monitoring System &
WINDUST pro Application Software - Handbook

Advertising