6.

SPECIFICATION

6.1

Instrument Specification

Measurement

Sensing technique:

Forward light scattering (12 to 20

o

) using

880 nm infra red source.

Measuring Range:

0.001 - 2,500 mg/m

3

over four ranges

0 - 2.5, 0 - 25, 0 - 250 and 0 - 2500 mg/m

3

.

Active range may be fixed or auto ranging.

Resolution:

0.001 mg/m

3

.

Zero stability:

Typically <2 µg/ m

3

/

o

C.

Span Stability:

±0.7% of reading /

o

C.

Operating Temp Range:

0 to 50

o

C (non condensing).

Storage Temp Range:

-20 to +55

o

C (non condensing).

Power Supply

Instrument Power Consumption:

Approx. 100 mA.

Battery:

4 x AA / MN1500 cells Alkaline or

rechargeable Nickel Cadmium.

Operating run time:

Alkaline Batteries (2700 mAh), typically

>20 hours,

Rechargeable NiCd cells (950 mAh),

typically >9 hours

Battery Charging

Internal NiCd fast charger circuitry.

Charge termination based on Negative

delta V detection with additional time-out

protection.

Charger must be disabled for dry cell

operation.

Charge Rate

Fast charge rate 450 mA,

Standby charge rate 55 mA approx.

Power adapter (-PC18)

Universal with input voltage range

100 - 240 VAC, 47 - 63 Hz.

Output 12 VDC at 0.8 A.

Supplied with connectors suitable for

USA, UK, European and Australian mains

outlet sockets.

The instrument may be powered

continuously from the external power

adapter.