4.

SAMPLING ACCESSORIES

When used in static monitoring applications, the sampling probe’s

measurement chamber should have some form of aspiration, either in the

form of an Apex (or Vortex) sampling pump or a fan aspirator unit.
Maintaining airflow through the probe eliminates potential errors due to

orientation of the probe inlet and size selectivity.
The gravimetric and respirable adapters incorporate 25 or 37 mm filter

cassettes and provide a convenient method of gravimetrically calibrating

MicroDust pro instruments.

4.1

Gravimetric Adapters

The gravimetric adapter units may be used in conjunction with a sampling

pump to enable all sampled particulate matter to be drawn through the probe

measurement chamber and then deposited on to a 25 or 37 mm filter. There

are three gravimetric adapter options available.

Total Suspended Particulate (TSP)

All sampled particulate is monitored in real time and then deposited

on to a sample filter. The design provides no size selective

characteristics.

Respirable (4 µm)

The respirable adapter design incorporates a Cyclone size selective

unit. Only the respirable fraction of particulate passes through the

probe and is deposited on to the filter.

PUF (Polyurethane Foam Filter)

The PUF gravimetric adapter uses foam filter inserts to provide size

selective sampling characteristics. Filters are available for PM

10

,

Respirable (4 µm) and PM

2.5

size fractions.

All components fit together using “O” rings for sealing. These should be

checked for signs of damage, which may result in erroneous readings due to

air leakage. The adapters enable standard gravimetric techniques to be used

and provide a cross check between the readings of the MicroDust pro and a

weighed sample.
If errors are found between the instruments reading and the gravimetric

value, a correction figure can be determined and applied to give a closer

correlation of the two readings. (Please refer to Calibrating for a Specific Dust

Type in Section 2.3.2.)
The sampling probe should be inserted into the adapter block and correctly

aligned over the sample inlet. The locating spigot of the Cyclone, TSP inlet

tube or PUF filter holder should be inserted into the probe’s sample inlet so

that the complete assembly cannot rotate or move laterally on the probe.

Page 43 of 60

Sampling

Accessories