Oxymitter 4000, Alarms o2 heater open, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 103

Advertising
background image

Instruction Manual

IM-106-340, Rev. 4.0
May 2006

8-11

Oxymitter 4000

Figure 8-7. Fault 5, Open Heater

Fault 5, Open Heater

Figure 8-7 shows the electronic assembly for an
Oxymitter 4000 with a membrane keypad (upper
view) and a Oxymitter 4000 with an LOI (lower
view).

Membrane Keypad

When Fault 5 is detected, the HEATER LED
flashes once, pauses for three seconds, and
repeats.

1. Remove power.
2. Remove the electronic assembly per

"Electronic Assembly Replacement" in
Section 9: Maintenance and Service.

3. Using a multimeter, measure the resistance

across the terminals of heater connector,
J8.

4. The measurement should be approxi-

mately 72 ohms. If the heater is open, see
"Heater Strut Replacement" in Section 9:
Maintenance and Service.

LOI

When Fault 5 is detected, the LOI displays the
"O2 Heater Open" message.

1. Remove power. Unscrew and remove the

LOI module from the electronic assembly.

2. Perform the diagnostic steps 2 through 4

shown for the membrane keypad.

37260023

LOI

KEYPAD

Alarms

O2 Heater Open

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON