Mechanical installation, Selecting location, Probe installation – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 40: Mechanical installation -2, Selecting location -2 probe installation -2, R to "selecting location" in section, Oxymitter 4000

Advertising
background image

Oxymitter 4000

2-2

Instruction Manual

IM-106-340, Rev. 4.0

May 2006

MECHANICAL
INSTALLATION

Selecting Location

1. The location of the Oxymitter 4000 in the stack or flue is most important

for maximum accuracy in the oxygen analyzing process. The Oxymitter
4000 must be positioned so the gas it measures is representative of the
process. Best results are normally obtained if the Oxymitter 4000 is
positioned near the center of the duct (40-60% insertion). Longer ducts
may require several Oxymitter 4000 units since the O

2

can vary due to

stratification. A point too near the wall of the duct, or the inside radius of
a bend, may not provide a representative sample because of the very
low flow conditions. The sensing point should be selected so the
process gas temperature falls within a range of 32° to 1300°F
(0° to 704°C). Figure 2-1 through Figure 2-8 provide mechanical
installation references. The ambient temperature of the integral
electronics housing must not exceed 185°F (85°C). For higher ambient
temperatures, we recommend the remote mounted electronics option.

2. Check the flue or stack for holes and air leakage. The presence of this

condition will substantially affect the accuracy of the oxygen reading.
Therefore, either make the necessary repairs or install the Oxymitter
4000 upstream of any leakage.

3. Ensure the area is clear of internal and external obstructions that will

interfere with installation and maintenance access to the membrane
keypad or LOI. Allow adequate clearance for removal of the Oxymitter
4000.

Probe Installation

1. Ensure all components are available to install the Oxymitter 4000. If

equipped with the optional ceramic diffusion element, ensure it is not
damaged.

2. The Oxymitter 4000 may be installed intact as it is received.

NOTE
An abrasive shield is recommended for high velocity particulates in the flue
stream (such as those in coal-fired boilers, kilns, and recovery boilers).
Vertical and horizontal brace clamps are provided for 9 ft and 12 ft (2,75 m
and 3,66 m) probes to provide mechanical support for the Oxymitter 4000.
Refer to Figure 2-6.

3. Weld or bolt adapter plate (Figure 2-5) onto the duct.
4. If using the optional ceramic diffusion element, the vee deflector must

be correctly oriented. Before inserting the Oxymitter 4000, check the
direction of gas flow in the duct. Orient the vee deflector so the apex
points upstream toward the flow (Figure 2-7). This may be done by
loosening the setscrews and rotating the vee deflector to the desired
position. Retighten the setscrews.

Do not allow the temperature of the Oxymitter 4000 electronics to exceed 185°F (85°C) or
damage to the unit may result.

Advertising