Oxymitter 4000, Alarms o2 t/c open, Instruction manual – Emerson Process Management OXYMITTER 4000 IM-106-340 User Manual

Page 99

Advertising
background image

Instruction Manual

IM-106-340, Rev. 4.0
May 2006

8-7

Oxymitter 4000

Figure 8-3. Fault 1, Open Thermocouple

Fault 1, Open Thermocouple

Figure 8-3 shows the electronic assembly for an
Oxymitter 4000 with a membrane keypad (upper
view) and a Oxymitter 4000 with an LOI (lower
view). The upper view also shows J1 and test
points TP1 through TP6, located on the
microprocessor board, below the membrane
keypad or the LOI module.

Membrane Keypad

When Fault 1 is detected, the HEATER T/C LED
flashes once, pauses for three seconds, and
repeats.

1. Check connector J1. Ensure the connector

is properly seated.

2. Using a multimeter, measure the voltage

from TP3+ to TP4-. If the reading is 1.2
VDC ±0.1 VDC, the thermocouple is open.

3. Remove power. Disconnect J1. Measure

the resistance across the red and yellow
thermocouple leads. The resistance should
be approximately 1 ohm.

4. If the thermocouple is open, see "Heater

Strut Replacement" in Section 9:
Maintenance and Service.

LOI

When Fault 1 is detected, the LOI displays the
"O2 T/C Open" message.

1. Remove power. Unscrew and remove the

LOI module from the electronic assembly.

2. Reconnect power to the Oxymitter 4000.
3. Perform the diagnostic steps 1 through 4

shown for the membrane keypad.

37260019

LOI

KEYPAD

DIAGNOSTIC

ALARMS

TEST

POINTS

HEATER T/C

HEATER

O2 CELL

CALIBRATION

CALIBRATION RECOMMENDED

O2 CELL mV +

O2 CELL mv -

HEATER T/C +

HEATER T/C -

INC

INC

DEC

DEC

HIGH

GAS

LOW

GAS

CAL

TEST GAS +

PROCESS -

% O2

SW2

TP1

J1

TP2

TP3

RED

YEL

GRN

ORG

TP4

TP5

TP6

ON

Alarms

O2 T/C Open

Advertising